Error codes

Prev Next

If an error is set, an error response is generated.

Examples

Query

Response

Meaning

MU 1

er 100

unknown command

IY K 500

er 101

unknown parameter

IY A 4000

er 102

parameter out of range

If an internal error is set, the response is integrated in the standard response:

TE

te 70-80,71,23-21,22;er 202,A

LED voltage at channel A out of range

List of error codes

Error code number

Description

100

Unknown command

101

Unknown parameter

102

Parameter out of range

103

Command is not allowed in the current operating mode

200

Supply voltage exceeds limits

201

Analog control voltage exceeds limits

202

LED Voltage out of range

203

Output voltage exceeds limits (Programmed in the Corona II module)

204

Warning temperature exceeded

205

Maximum temperature exceeded

206

Shutdown active

210

Write error to EEPROM in Driver Unit

211

Read error from EEPROM in Driver Unit

212

Write error to EEPROM in Corona

213

Read error from EEPROM in Corona

214

Fan failure detected at ACTRL

215

Programmable voltage value out of range (Programmed in the XLC4 control unit)