X6 PWM interface / external switch
    • 04 Jul 2024
    • 2 Minutes to read
    • PDF

    X6 PWM interface / external switch

    • PDF

    Article summary

    The following connector is required:

    • Würth Elektronik 691 325 110 008

    The following counterpart is required:

    • Würth Elektronik 691 364 100 008

    control unit XLC4-1 and XLC4-1A

    pin

    name

    description

    1

    ON

    IN

    switch signal for PWM mode operation

    • switches on the lighting in PWM-mode

    2

    PWC

    IN

    control signal for PWM mode operation

    • controls the LED current in PWM-mode

    3

    ENC

    IN

    do not use

    4

    FAIL

    OUT

    failure signal

    • signals error states in PWM-mode, Bulb-mode and Analog-mode

    5

    STAT

    OUT

    do not use

    6

    COMMON

    GND

    common ground for pins 1 – 5

    7

    SW1

    switch

    8

    SW2

    switch

    The signal ON switches the Corona II module to PWM-controlled mode.
    The duty cycle of signal PWC controls the LED current in PWM-controlled mode.
    The signal FAIL sets error states in PWC-controlled mode, Bulb mode and Analog mode. Refer to PWM-controlled mode for further information.

    The two inputs ON and PWC are opto-coupled inputs. They both use the ground line COMMON.
    The output signal FAIL is opto-coupled with an open collector circuit to the ground line COMMON.

    control unit XLC4-1F4

    pin

    name

    description

    1

    ON_1

    IN

    external LED control channel A

    2

    ON_2

    IN

    external LED control channel B

    3

    ON_3

    IN

    external LED control channel C

    4

    FAIL

    OUT

    failure signal

    5

    STAT

    OUT

    do not use

    6

    COMMON

    GND

    common ground for pins 1 – 7

    7

    ON_4

    external LED control channel D

    Signal ON - input characteristics

    control unit XLC4-1

    ON Max voltage

    Input voltage ON ≤ 12 V

    ON Max current

    Input current ON ≤ 15 mA

    ON Low current

    Input current ON < 0.2 mA

    ON High current

    Input current ON > 0.7 mA (Recommended > 2 mA)

    control unit XLC4-1A, XLC4-1F4

    ON Max voltage

    Input voltage ON ≤ 25 V

    ON voltage

    ≥ 2.5 V

    OFF voltage

    ≤ 0.7 V

    Signal PWC - input characteristics

    PWC Max voltage

    Input voltage PWC ≤ 6 V

    PWC Max current

    Input current PWC < 15 mA

    PWC Low current

    Input current PWC ≤ 0.7 mA

    PWC High current

    Input current PWC > 1.3 mA (Recommended > 5 mA)

    If PWC needs higher voltages use a series resistor.

    Signal FAIL - output characteristics

    FAIL Max voltage

    Output voltage FAIL ≤ 12 V

    FAIL Max current

    Input current FAIL ≤ 15 mA (Recommended 5 – 10 mA)

    NOTE

    The output transistor of FAIL is switched on if no error is set. Use a pull-up resistor in the receiver. An interrupt at the FAIL signal line causes an error (FAIL = H) at the receiver.

    Sample circuit

    NOTE - resistor value

    The resistor value refers to the supply voltage of 24 V. Calculate the value of the resistor dependent to the supply voltage of the power supply.

    Signal SW1, SW2

    SW1 and SW2 ports are used to activate an external cooling device. When the programmed level is reached, the connection between SW1 and SW2 is closed.

    • The maximum current within the internal switch is limited to 1.5 A

    • The voltage at SW1 and SW2 must be lower than the supply voltage

    • The voltage at SW1 must be higher than the voltage at SW2


    Was this article helpful?