    Status LEDPossible causeDevice error codeAction
    It does not turn green immediately after switching on the power supply.The fuse has been tripped due to an incorrect input voltage.
    1. Switch off the power supply.
    2. Contact service.
    YellowThe internal temperature is above the defined warning limit.DEV_CTRL_WARNING_
    TEMPERATURE_TOO_HIGH
    Decrease ambient temperature and improve cooling.
    RedThe internal temperature has reached the defined error limit.DEV_CTRL_ERROR_
    TEMPERATURE_TOO_HIGH
    The camera automatically switches to safety mode and indicates an image with a pin stripe test pattern on a black background.

    1. Switch off the power supply and let the camera cool down.
    2. Check the ambient conditions (0 °C – 60 °C; 32 °F – 140 °F) and improve cooling.
    3. Switch on the power supply.