Command VT

Prev Next

Command VT is used to read the values of the supply voltage.

Structure

Command

Notes

CommandID

VT

VT works in all XLC4 operating modes

Parameter

None

Parameter

Value

Description

None

The current values of the supply voltage are queried

Response

vt, space, input , analog , channel 1, channel 2, channel 3, channel 4

General notes

The output values are in mV. For the voltage of the LEDs apply a factor of 18,46.

Example

Query

Response

Action

VT

vt 24758,0,1347,1343,1350,1350

Displays the current values:

Supply voltage: 24,758 mV

Analog input voltage: 0 mV

LED voltage at channel 1: 1,347 mV × 18.46 = 24,866 mV

LED voltage at channel 2: 1,343 mV × 18.46 = 24,792 mV

LED voltage at channel 3: 1,350 mV × 18.46 = 24,921 mV

LED voltage at channel 4: 1,350 mV × 18.46 = 24,921 mV

SV 22000,25500

sv 22000,25500

The new values are set:

  • 22000 mV mimimum level

  • 25500 mV maximum level