Command TE is used to query the values of the internal temperature sensors of the XLC4 control unit and the Corona II modules.
Structure
Command | Notes | |
---|---|---|
CommandID | TE | TE works in all XLC4 operating modes |
Parameter | None | |
Parameter | Value | Description |
None | ||
Response | ||
te , space , tempPair , “,” , tempPair , “,” , tempPair, “,” , tempPair | ||
tempPair = tval , [“-” , tval] | ||
General notes | ||
The value list in the response message contains the temperature values of the four output channels of the XLC4 (channel A first). If Corona II illumination modules are connected, the internal temperature values of the Coronas are added to the corresponding channel temperature values (separated by -).
Example
TE
te 23-22 , 22 , 23-21 , 22
Note that only the first channel of a Corona II module is connected to a temperature sensor. The channels with a temperature pair (“-” separated) are connected to the first channel of a Corona II module.
In this example, there are 2 Corona II modules connected:
One to channel A and B
One to channel C
The TE command can be used to determine the number of connected Corona II modules.